Nashville
Surgical Instruments
The Kumar PRE-VIEW* Cholangiography
Kumar
PRE-VIEW* Cholangiography Clamp & Catheter (Latex Free)
INDICATIONS
FOR USE:
The Kumar PRE-VIEW* Cholangiography Clamp,
the (Kumar) Clamp and the Kumar PRE-VIEW* Cholangiography Catheter
(the Catheter) are intended to provide a method of laparoscopic
cholangiography, and are to be used strictly with each other.
(US Patent No. 5,224,931)
CONTRAINDICATIONS:
The device is not intended for use when the
associated surgical techniques are contraindicated.
DEVICE DESCRIPTION:
The Kumar PRE-VIEW* Cholangiography Clamp is a reusable, stainless-steel
5mm laparoscopic grasper with long (4 cm.) atraumatic jaws and a
channel for introduction of the Kumar PRE-VIEW* Cholangiography
Catheter. A valve at the top of the channel can be opened to insert
the Catheter or closed to maintain pneumo-peritoneum.
The Catheter is a translucent polyurethane tubing of 16 ga. and
carries a 19 ga., 1.25 cm. long needle at the end.
The Clamp applies completely across the lower part of the body of
the gallbladder and the catheter needle punctures the Hartmann's
pouch of the gallbladder for biliary access or aspiration.
FOR CYSTIC DUCT OBSTRUCTION OR HYDROPS: see below and under
Acute Cholecystitis.
Instructions for Use:
(This is not a primer of surgical techniques or cholecystectomy)
1.
Apply the Kumar PRE-VIEW* Cholangiography Clamp through the right
mid-subcostal port at the beginning of the laparoscopic cholecystectomy
procedure. Use it as a regular grasper for traction of the gallbladder
during cystic duct dissection.
2. When cystic duct dissection is complete, milk cystic duct
towards the gallbladder to eliminate sludge, mucous or gravel from
the cystic duct. This enhances cystic duct patency.
3. Re-position and apply the Kumar Clamp completely across
the lower part of the body of the gallbladder (Fig. 1), just above
the Hartmann's pouch. During this step, it is helpful to grasp the
Hartmann's pouch with a grasper through the epigastric port and feed
it into the jaws of the Clamp. (Usually, the tips of the Clamp jaws,
at this time, point at the cystic artery, Fig.1). Caution: Avoid injury
to the lower surface of liver from the tips of the Clamp jaws.
4.
By sterile technique, remove Catheter from package. Remove needle
guard & flush Catheter.
5. Open the Clamp channel valve. Insert and advance Catheter
into the Clamp channel.
6. Laparoscopically visualize the Catheter needle as it exits
the lower end of the channel. Flush Catheter again to eliminate any
air bubbles.
7. Advance the Catheter and puncture Hartmann's Pouch in a
Bull's eye manner, adjusting the Clamp by tilting or rotating as needed
(Figure 1). Rarely it is necessary to guide the Catheter / needle
with a grasper through the epigastric port.
Figure 1
8. Aspirate to confirm biliary
access. Inject dye for cholangiography. Fluoroscopy is preferred.
9.
After completion of cholangiography, open Clamp jaws and use Catheter
to aspirate and empty the gallbladder. The emptied gallbladder is
much easier to separate from the liver bed and extract from the portsite.
10. Remove Catheter under direct vision. Close channel valve
(or leave it open as a smoke vent!) and use the Kumar Clamp as a grasper
to finish the operation.
Cystic Duct Obstruction due to small stones, gravel,
sludge or mucous (without Hydrops or acute cholecystitis) can be resolved
by milking the cystic duct towards the gallbladder. Dye can flow around
a tiny stone in the cystic duct without flushing the stone into the
common bile duct. The chances of flushing such a stone are no different
than during cystic duct cannulation. The clinical experience is that
such tiny stones do pass uneventfully.
(Aspiration of Hydrops)
|
Acute Cholecystitis (Hydrops) due to a stone that is
impacted at the neck of the gallbladder:
1. The gallbladder is usually so distended that it cannot be grasped.
Introduce a 5 mm. grasper through the right lateral port and, with
the jaws open, push the body/fundus of the gallbladder towards the
diaphragm as far as possible (Figure 2 above).
2. Introduce the Kumar Clamp through the right mid-subcostal port.
Advance the Clamp jaws to the lowest part of the body of the gallbladder,
keeping the jaws open (Figure 2 above).
If visualization of the cystic duct area is limited, it is helpful
to add an additional port through which a fan-shaped retractor can
be introduced to retract the viscera and improve visualization.
3. Advance Catheter and aspirate gallbladder at the lowest, dependant
location.
4. Aspiration relieves the pressure that is pushing the stone into
the neck of the gallbladder. This allows the impacted stone to dislodge
and fall back into the gallbladder. Compression of the impacted stone
with a grasper through the epigastric port may also be helpful.
5. The decompressed gallbladder can now be grasped and cholangiography
dye can be injected, preferably through the same access or the needle
can be moved to a lower location.
WARNINGS
AND PRECAUTIONS
1.
The device is for use by qualified surgeons, who are familiar with
this device prior to surgery.
2. Always operate the instruments under direct laparoscopic
monitoring.
HOW
SUPPLIED
The Kumar PRE-VIEW* Cholangiography Clamp is a reusable 5 mm
grasper, supplied non-sterile.
The Kumar PRE-VIEW* Cholangiography Catheter is 76 cm. long,
16 ga. with a 19 ga., 1.25 cm. long needle. It is supplied sterile
and is for single patient use only. Discard after use. DO NOT RESTERILIZE
OR REUSE.
STERILIZATION
The Kumar PRE-VIEW* Cholangiography Clamp: Clean & steam autoclave
in accordance with the guidelines of Association for Advancement of
Medical Instruments (AAMI): Standards and Recommended Practices: Sterilization
in Healthcare Facilities. CLEAN & STERILIZE AFTER EACH USE.
Additional Instructions are provided in each Clamp package and below.
The Kumar PRE-VIEW* Cholangiography Catheter is supplied sterile
and is for single patient use only. .Sterility is guaranteed unless
package is opened or damaged. DO NOT RESTERILIZE.
CAUTION: Federal (USA) law restricts this device to use
by or on the order of a physician.
Manufactured for: Nashville Surgical Instruments, 2005 Kumar Lane,
Springfield, TN, 37172 USA
*U.S. Patent No. 5,224,931 and *Trademark.
INSTRUCTIONS FOR
CLEANING AND STERILIZATION
For Kumar PRE-VIEW* Cholangiography Clamp Only
Preparation
Inspect
instrument after each use for damaged or loose parts** and proper
function. Remove any gross contamination. To assure cleaning of all
surfaces, the clamp jaws, channel valve and the flush-port cap must
be open.
Pre-Rinse
1.
An initial cold water and blood/protein-dissolving enzyme solution
rinse (or soak for heavy contamination) helps remove blood, tissue
and debris from device lumens, joints & serrations.
2.
Flush and clean interior of instrument through Flush Port (after undoing
the black plastic cap), and the catheter channel (with the metal one-way
valve at the top) with cold water and blood/protein-dissolving enzyme
and antiseptic solution. Allow solution contact for 5minutes.
3.
Pre-clean clamp jaws and tips with a brush.
Washing / Decontamination
1.
Place instrument in a separate wire basket on top shelf of the automated
washer / decontaminator and allow a full cycle run.
2.
Allow instrument to air dry in basket. Then remove & inspect for
any residual contamination.
3. Allow residue of rinse
water to remain in instrument channels. This allows steam to form
during the sterilization process. If needed, inject distilled or de-ionized
water in to channels.
Packaging
Follow
institutional Policy and Procedures. Wrap properly for steam sterilization
with sterilization indicator strip in lowest and most inaccessible
portion of the packaging. Use two layers of non woven disposable sterilization
wrap on outside of packaging. Seal with sterilization indicator tape.
Label and initial.
Sterilization
l.
Follow sterilizer manufacturer instructions for use of the steam sterilizer.
2.
Recommended Minimum Exposure Times (Minutes after conditioning) for
Steam Sterilization: Gravity 250°F (121° C): 30
minutes, Gravity 270°-274°F (132°-134°C):
15 minutes, Pre-vac 270°-274°F
(132°- 134°C): 4 minutes.
3. Remove from sterilizer and allow to cool on sterilizer
rack. Assure package integrity
and store on designated shelves for sterile instruments.
4. Inspect instrument closely before each use for
damaged, loose parts and proper function.
Loose
Parts Advisory
There are three (3) loose or removable parts on the
Clamp: the black plastic cap over the flush port, the channel valve
assembly and the screw at the bottom that holds the valve in place.
None of these enter the body during normal use. Caution is necessary
when the operation is converted to open surgery.
* Patent and Trademark